Epson Powerspan Especificaciones Pagina 100

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 515
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 99
ZVx INITiate Subsystem
1043.0009.50 3.61 E-15
3.6.8 INITiate Subsystem
The INITiate subsystem controls the initialization of the trigger subsystem.
COMMAND PARAMETER UNIT COMMENT
INITiate
:CONTinuous
[:IMMediate]
<boolean>
--
--
-- no query
INITiate:CONTinuous
This command determines if the trigger system is continuously initiated ("Free Run")..
Syntax: INITiate:CONTinuous ON | OFF
Example: "INIT:CONT OFF"
Features: *RST value: ON
SCPI: conforming
Setting "INITiate:CONTinuous ON" corresponds to the CONTINUOUS SWEEP, i.e. the sweep is
cyclically repeated. The setting "INITiate:CONTinuous OFF" corresponds to the function SINGLE
SWEEP.
INITiate[:IMMediate]
The command initiates a new sweep or starts a single sweep.
Syntax: INITiate[:IMMediate]
Example: "INIT"
Features: *RST value: -
SCPI: conforming
This command is an event, which is why it is not assigned an *RST value and has no query.
Vista de pagina 99
1 2 ... 95 96 97 98 99 100 101 102 103 104 105 ... 514 515

Comentarios a estos manuales

Sin comentarios