ZVx CALCulate Subsystem
1043.0009.50 3.23 E-15
3.6.3.3 CALCulate:GDAPerture Subsystem
The CALCulate:GDAPerture - subsystem defines the parameters for the group delay and the aperture.
COMMAND PARAMETER UNIT COMMENT
CALCulate<1..4>
GDAPerture
:MODE
[:SPAN]
:SCOunt
STEP | FREQuency
<numeric_value>
<numeric_value>
HZ
CALCulate[1...4]:GDAPerture:MODE
This command switches between the aperture defined as a number of points (STEP) or as a fixed
frequency value..
Syntax: CALCulate[1...4]:GDAPerture:MODE
Example: "CALC:GDAP:MODE STEP"
Features: *RST value: STEP
SCPI: device-specific
CALCulate[1...4]:GDAPerture[:SPAN]
This command defines the aperture as a fixed frequency value..
Syntax: CALCulate[1...4]:GDAPerture[:SPAN] <numeric_value>
Example: "CALC:GDAP 0.5"
Features: *RST value: -
SCPI: conforming
CALCulate[1...4]:GDAPerture:SCOunt
This command defines the aperture as a number of points..
Syntax: CALCulate[1...4]:GDAPerture:SCOunt <numeric_value>
Example: "CALC:GDAP:SCO 12"
Features: *RST value: 10
SCPI: device-specific