Example of Measurement Program
100 CMT “NEGATIVE IMPEADANCE”
110 DISP ‘START FREQUENCY”
120 START=10E6
130 DISP “STOP FREQUENCY”
140 STOP=100E6
150 SWT2
160 RBW=3 KHz
170 !****S PARA
180 FNC4;SPI3
190 DISP “S11”
200 SWTRG
210 E=MA;F=MB
220 FNC5;GPP3
230 DISP “S21”
240 SWTRG
250 I=MA;J=MB
260 FNC6;GPP3
270 DISP “S12”
280 SWTRG
290 G=MA;H=MB
300 FNC7;SPI3;DPB1
310 DISP “S22”
320 SWTRG
330 RA=MA;RB=MB
340 !
350 !****N Z CAL
360 BEEP
370 DISP “CSL”
380 WAIT 1
390 DSP1
400 <RC,RD>=(<1,0>+<E,F>)*(<1,0>-<RA,RB>)+(<1,0>+<RA,RB>)*(<1,0>-<E,F>)
410 <RE,RD>=<RC,RD>-<2,0>*<G,H>-<2,0>*<I,J>+<2,0>*<G,H>*I,J>
420 <RE,RF>=<RE,RF>/((<1,0>-<E,F>)*(<1,0>-<RA,RB>)-<G,H>*<I,J>)
430 FNC3;IMP2;DPB1
440 A=RE*50;B=RF*50
450 SCL1;DIV=200;REF=0
460 SCL2;DIV=200;REF=0
470 END