7-29
4
To change the default condition, press the
OPTION
key. NO will
change to YES on the display and you will be able to select a
different tube type when you start a run (see Chapter 5 for more
information).
5
Press
ENTER
.
The following display appears:
This display lets you change the sample tube time constant. There
are two default values for sample tube time constants on this
display: one for sample tubes with a reaction volume of 100
µ
L, and
one for sample tubes with a reaction volume of 50
µ
L. There is a
time constant default value for sample tubes with a reaction volume
of 20
µ
L that appears on a separate display. Press
STEP
to view
this display:
6
To change any of these time constant values, press the appropriate
numeric keys, then press
ENTER
. To move the cursor from one
time constant value to the next, press
ENTER
. To move from the
first display to the second, press
STEP
.
7
Press
STOP
to return to the first Configuration display.
Rxn vol=100uL T= 9.5
Rxn vol= 50uL T= 7.0
Rxn vol= 20uL T= 5.0