8-8
When the clock reaches three minutes, the following display
appears:
11
Measure the temperature of well E1 using the digital thermometer
and record this temperature as T(40). See Figure 8-3 on the next
page.
12
Press
ENTER
.
The temperature of the sample block and heated cover will go
to 95 degrees C, and the following display will appear:
This display shows the current temperatures of the block cover
(Cvr= xxC) and the sample block (Blk = xx.xC).
When the temperature of the block cover is within ten degrees of
the sample block temperature, the following display appears:
This displays shows the current sample block temperature
(Blk=95.0C) and a clock, which counts up from zero in minutes
and seconds (Time=MM:SS).
Record Temperature
Time=MM:SS Blk=40.0C
Going to 95°C...
Cvr= xxC Blk = xx.xC
Wait 3 minutes
Time=MM:SS Blk=95.0C