Epson C82302/04 (Serial I/F) Manual de usuario Pagina 229

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 294
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 228
7-41
Figure 7-9 shows an example status record display for a
successful run.
Figure 7-9
,A typical status record display.
The example in the figure tells you that CYCL program #2, which
is one of the linked programs specified in Method #7, was
aborted at 14:47 (2:47 PM) on December 21, 1990 and that this
status record is Record #5 of your entire run.
Power Failure xxx.x hrs This message is also a fatal error
message and will appear if a power failure
occurred during the run. It shows how long
the power was off.
Abort mm/dd/yy hh:mm This message will appear if you pressed
the
STOP
key during the run. It shows the
time and date the run was aborted.
Pause Timeout hh:mm The pause timeout limit was exceeded
during the run.
Fatal Setpoint Error The sample block did not reach its
setpoint in a reasonable period of time.
Pause xx:xx at xx.xC The run for this time and at this
temperature.
CYCL #2/7 5
Abort 12/21/90 14:47
Program
Type
Program #/Method #
Record #
Status Message
Date of Abort
Time of Abort