Epson C82302/04 (Serial I/F) Manual de usuario Pagina 277

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 294
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 276
10-3
Sample Temperature Range
The temperature range of the sample block is:
4.0 degrees to 99.9 degrees C.
Sample Temperature Accuracy
The displayed sample temperature matches the average true
sample temperature within +/0.75 degrees C.
Displayed Sample Temperature
The displayed sample temperature will match the specified
setpoint temperature to within +/0.2 degrees C, centered around
the setpoint temperature two minutes after the sample
temperature is within 1 degree C of setpoint.
Sample Block Temperature Uniformity
The sample block temperature uniformity is:
+/0.5 degrees C 30 seconds after the clock starts, over the
range 50.0 to 99.9 degrees C.
Sample Temperature Overshoot
An average of 96 sample temperatures shows less than 0.5
degrees C overshoot for upramps and downramps.
Figure 10-1 illustrates block temperature overshoot after ramping
up.
Vista de pagina 276
1 2 ... 272 273 274 275 276 277 278 279 280 281 282 ... 293 294

Comentarios a estos manuales

Sin comentarios