8-14
Figure 8-5
The probe assembly in sample wells A1 and A2.
5
Slide the heated cover forward and turn the cover knob clockwise
until the white mark on the cover knob is aligned with the white
mark on the cover.
6
Turn on the digital thermometer by moving the ON-OFF/RANGE
switch to the 200 position.
7
Turn on the GeneAmp PCR System 9600 and set up a
two-temperature CYCL program with the following
parameters:
Setpoint #1 Temperature = 95°C
Hold Time = 2:00 minutes
Ramp Time = 0:00 minutes
Setpoint #2 Temperature = 40°C
Hold Time = 2:00 minutes
Ramp Time = 0:00 minutes
Cycles = 99
Note
Refer to Chapter 4 for detailed instructions on how to set up
and run a CYCL program.