  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 256
8-7
9
Press the
OPTION
key twice to move the cursor to DIAG, then
press
ENTER
. The following display appears:
10
Run the Verify Calibration Diagnostic Test (Test #5) by pressing
5
,
then
ENTER
.
The temperature of the sample block and heated cover will go
to 40 degrees C, and the following display will appear:
This display shows the current temperature of the heated cover
(Cvr= xxC) and sample block
(Blk = xx.xC).
When the temperature of the heated cover is within ten
degrees of the sample block temperature, the following
display appears:
This displays shows the current sample block temperature
(Blk=40.0C) and a clock, which counts up from zero in minutes
and seconds (Time=MM:SS).
Select function
D
IR-CONFIG-DIAG-DEL
Enter Diag Test #1
REVIEW HISTORY FILE
Going to 40 ºC...
Cvr= xxC Blk = xx.xC
Wait 3 minutes...
Time=MM:SS Blk=40.0C